How does FCS Express implement software compensation?

The emission spectrum (the wavelengths of light at which a given fluorophore emits) generally overlaps several detection channels of a typical commercial flow cytometer. Although most of the fluorescence signal will be recorded in a single primary channel, it is common that signal from the same fluorophore can be detected in two (or sometimes more) secondary channels. Compensation is the process of removing the unwanted fluorophore signal from secondary channels. A classic example is that although the fluorescence from fluorescein is primarily measured in the “green” channel (~530 nm), a significant amount of fluorescein fluorescence can also be detected in the “yellow” channel (~570 nm). Thus, if you are trying to detect another fluorochrome in the “yellow” channel (i.e. PE), the signal you observe in the “yellow” channel will comprise fluorescence from both PE and FITC

In many flow cytometers, compensation is built directly into electronics of the flow cytometer. This is typically called hardware compensation or analog compensation. There would be dials or buttons that would allow the user to change the compensation parameters before acquiring the data.

The dial would typically be labeled something like:. FL2-%FL1, or PE-A-% FITC-A. Where FL2 is the total amount of fluorescence observed or measured in the FL2 channel and the %FL1 is the percentage of fluorescence measured in the FL2 channel that is from the FL1 fluorochrome. By changing the dial or adjusting the setting, the user is changing the amount of fluorescence from the FL1 or FITC parameter to subtract from the total fluorescence measured in the FL2 or PE parameter. For example, if the FL2-%FL1 or PE-A-% FITC-A value was 20, this means that the true FL2 value is the observed FL2 value minus 20% of the observed FL1 value.

Previous How do I use R Integration with FCS Express?
Next If my data does not have a Time parameter, can I create one?